Manuscript Offprint
arXiv:2610.05839Dual-Rate Force-Image Control with Model-Based Orientation Limits for Robotic Ultrasound
Tyler Foster , Qiang Zhang , Member, IEEE, A B M Tahidul Haque , and Anh Thu Nguyen
Abstract—Robotic ultrasound couples a high-rate contact-force loop with slower, delayed image feedback, so image-guided ultrasound probe rotation can perturb contact force before the resulting image response is observed. We derive a closed-form orientation-rate limit that bounds the modeled rotation-induced estimated-force excursion over a finite horizon while accounting for disturbance rejection by the fast force loop. The limit depends on local contact stiffness, force-loop gains, a conservative rotation-to-force gain bound, the excursion budget, and the prediction horizon. We implement this model in a dual-rate controller with timestamp-based delay reconstruction and joint-torque-based force estimation, and evaluate it on a curved gelatin phantom using paired controller comparisons and component ablations. Relative to unconstrained image guidance, the proposed rate-limited controller reduced first-second root-mean-square (RMS) estimated-force error by 0.40 N while increasing cue-convergence time by 0.94 s. A fixed rate cap near the analytically predicted ceiling produced no resolvable difference in force error and converged 0.32 s faster, indicating that the principal practical value of the model is the rate-design rule rather than online prediction. Delay reconstruction had no resolvable effect at the tested latency. A single-subject popliteal scan demonstrated feasibility, although the image cue was noise-limited on heterogeneous tissue.
I. INTRODUCTION
Brightness mode (B-mode) ultrasound provides real-time, radiation-free imaging at comparatively low cost, but its image quality depends strongly on how the probe interacts with tissue. A sonographer continuously coordinates probe position, orientation, and contact force while responding to surface geometry, tissue compliance, and the image itself; robotic ultrasound is therefore a coupled sensing-and-control problem rather than a trajectory-following task. Reviews accordingly identify scan-path generation, probe-tissue force regulation, and image-quality or orientation optimization as the central components of autonomous systems [1]-[3], and recent surveys [4], [5] note open gaps, among them that orientation-related interaction forces are often simplified and that real-time image feedback is still rarely integrated into online scan optimization [5].
Image features extracted directly from ultrasound can close the robot-positioning loop [6], and visual servoing on ultrasound confidence has been used to improve image quality through probe orientation [7]. Interaction-force information [8], [9] and probe-mounted distance sensing [10] have been used to estimate the local surface normal and regulate orientation, explicit force/moment control improves interaction under heterogeneous contact [11], ultrasound feedback has enabled autonomous screening of tubular structures [12], and image-quality estimates have been used to adapt the commanded force [13], [14]. Image and force sensing are therefore established as complementary. Learning-based methods have extended autonomy to navigation, compliant scanning, and complete clinical workflows [15]–[17], including policies trained under explicit force constraints [18]; this work instead keeps a physically structured force loop and uses the image only as a measured cue.
A control issue that is less explicitly addressed is the multirate dynamic coupling between force regulation and image-guided probe correction. Robot state and wrench estimates update at high rates, whereas B-mode images are acquired, transferred, and processed at much lower rates and with non-negligible latency. The difficulty is more fundamental than asynchronous sensing alone, because the two loops are coupled through the contact. An orientation correction computed from an image changes probe indentation, the contact moment, and consequently the normal force before a new image becomes available. Conversely, a fast force regulator reacts to the normal displacement induced by the orientation update and can partially oppose the intended correction. Dual-loop approaches have demonstrated image-dependent force adaptation and robust force tracking [14], Li et al. [19] incorporate trajectory-dependent feedforward and delay compensation for contact-force tracking, and Qian et al. [20] combine image guidance with biomechanics-informed force measurement. These approaches do not treat the image-guided orientation loop and the high-rate force loop as a delayed multirate system whose cross-coupling is predicted and used in the controller.
This paper treats that coupling as the control problem. We propose a dual-rate orientation controller in which the slow image-guided loop accounts for both its own latency and its effect on the fast contact loop: the image cue is reconstructed to the present instant from the frame's capture timestamp; the force excursion that a candidate correction would produce is predicted from a conservative bound on the rotation-to-force gain; and the prediction credits the fast loop with the portion of that excursion it will reject within the prediction horizon. This credit separates the method from a one-step safety filter, which must assume the excursion persists and therefore admits a lower rotation rate. While the correction demand is saturated, the constraint reduces to a ceiling on the rotation rate that depends only on the contact stiffness,
Tyler Foster, Qiang Zhang, and A B M Tahidul Haque are with the Department of Mechanical Engineering, The University of Alabama, Tuscaloosa, AL 35487, USA. Qiang Zhang is also with the Department of Chemical & Biological Engineering, The University of Alabama, Tuscaloosa, AL 35487, USA. Anh Thu Nguyen is with the Department of Computer Science, The University of Alabama, Tuscaloosa, AL 35487, USA. Corresponding author: Qiang Zhang, qiang.zhang@ua.edu.
the force-loop gains, the gain bound, the force budget, and the prediction horizon, so it can be evaluated before image-guided correction begins, once the local stiffness and a valid upper bound on rotation-to-force gain have been established. The platform has no force/torque sensor; contact force is estimated from joint torques. This raises the force noise floor and introduces model-dependent bias, so the force margins the controller must respect are small relative to the estimate's error. The contributions are:
-
a closed-form design rule for the maximum image-guided orientation rate that limits the modeled rotation-induced force excursion over a finite horizon while accounting for force-loop rejection;
-
a paired hardware evaluation on a curved phantom against force-only, newest-frame, and fixed-rate-cap baselines, with ablations isolating the constraint, the fast-loop credit, delay reconstruction, and online sensitivity estimation, showing that the ceiling sets the trade-off between cue-convergence time and force error, and that a fixed cap near the predicted rate matches the online constraint in force error; and
-
a single-subject popliteal scan testing feasibility on living tissue.
A. Platform
The platform (Fig. 1) is a Kinova Gen3 six-degree-of-freedom manipulator carrying a Telemed LF9-5N60-A3 linear probe in a rigid printed holder. No force/torque sensor is present; the contact wrench is estimated from the torque sensors in each actuator. This removes the cost, cabling, and sterilization burden of a wrist-mounted sensor at the expense of a higher noise floor and a model-dependent bias.
The arm is commanded in joint position at kHz through the manufacturer's low-level cyclic interface. A Cartesian velocity command is mapped to joint rates through a damped least-squares inverse of the geometric Jacobian and integrated to the streamed position command, as
with the measured joint coordinates and . B-mode frames are captured from the scanner display at Hz, and each is tagged with its capture time. The scanner's internal acquisition-to-display delay was not resolved and is not compensated.
B. Wrench Estimation from Joint Torques
With the measured joint torques and the probe's wrench on the tissue, the rigid-body model is
where is the gravity torque and is a joint-level residual. The gravity torque is computed by a quasi-static recursive Newton–Euler pass [21] with link parameters from the manufacturer’s model, the last link’s mass and center of mass updated for the probe and holder, and the

Fig. 1. Experimental setup: a Kinova Gen3 six-degree-of-freedom arm holds a Telemed LF9-5N60-A3 linear probe in a printed holder against the gelatin phantom; the host machine runs the controller and shows the live B-mode stream and its confidence map.
gravity vector corrected to its measured value at the base. The acceleration- and velocity-dependent (inertial and Coriolis) terms are omitted. Evaluated on the logged trajectories with the full model, they contribute at most 1.2 N to the regulated force in the first second after rotation onset and 0.8 N during the traverse for joint motion below 5 Hz (2.0 and 2.1 N below 10 Hz), and do not change the sign or significance of any controller comparison (Section VI). Including them online would require differentiating the velocity feedback, itself computed by the actuators; on the logged data this inflates the computed terms at least fivefold with noise.
The residual depends on joint position and on the direction of recent motion. It is modeled per joint by two branches, each a constant plus low-frequency harmonics of joint angle (periods of 45–290°) identified offline from free-space pose sweeps. The weight between branches evolves as a single stop (elastic–plastic) operator of width 1° [22], so the model blends linearly between branches over the first degree of travel after a reversal; it is a low-order model with no minor-loop memory. The estimated contact wrench is
where is after a per-joint second-order Butterworth filter with a 25 Hz cutoff, denotes the pseudoinverse, and is the residual model.
The regulated variable is the beam-axis component of the estimated contact force in a probe frame whose origin is the acoustic face,
where is the force component of , the base-to-probe rotation, a free-space tare of acquired while the arm holds position immediately before contact, and selects the beam axis, which points out of the face, so compression is positive. Because absorbs whatever residual the model leaves at that pose, it is preceded by a conditioning motion that walks each joint against and then along its press direction, so the tare is taken on the same branch that the subsequent press will traverse.

(a)
(b)
Fig. 2. Image cue on the phantom. (a) B-mode frame; the right side of the image is darker. (b) Its confidence map (blue 0 to red 1), with the depth at which confidence falls below (white) and its least-squares line (dashed). The confidence threshold is reached at shallower depth on the right, so falls with and, by (6), . Upper half of the image shown.
C. Achievable Accuracy
Since (3) has no independent force reference during operation, its accuracy bounds every downstream claim, and all force results are stated in estimate space. Static loading against a reference scale was repeated at eight arm poses, each holding five levels between 5 and 20 N under force control; the loop held every level within 0.02 N of its setpoint. Over the 40 contact dwells, the error of (3) against the scale was 0.41 N root-mean-square (RMS) with the residual model and 0.49 N without it, with maxima of 1.11 and 1.27 N; the model lowered the error at 30 of the 40 dwells. The mean error at each level lay between -0.02 and -0.28 N, but its size and sign varied with pose (from -0.72 to +0.48 N at 5 N), so a single scale factor would not remove it. After the joint reversal at retraction, the free-space reading was offset by 1.19 N RMS with the model and 1.45 N without it, the model over-correcting the reversal (mean +1.08 N against -1.41 N) but still reducing its magnitude. These figures bound the estimate rather than calibrate it.
III. ORIENTATION SENSING FROM ULTRASOUND
Misalignment with the surface or partial loss of acoustic coupling lowers confidence on one side of the aperture, where image quality falls off from the transducer face downward. We quantify this asymmetry with a dimensionless scalar cue I, with I = 0 for a laterally symmetric image.
The cue is computed from the ultrasound confidence map of Karamalis et al. [23], which assigns to each pixel the probability that a random walker released there reaches the transducer before the far field, with edge weights derived from intensity gradients under an attenuation model. Writing for the confidence at row and column , the map is obtained by solving a sparse Dirichlet problem on the pixel graph, here at on a fixed crop of the displayed image by direct factorization with the symbolic analysis performed once. The per-frame cost on the experimental hardware, including crop, resize, solve, and feature extraction, was ms (95th percentile 30.7 ms; 2% of frames exceeded the 33 ms frame interval). Because the map is solved in parallel with capture, the control signal always comes from the previous frame, or from two frames back when a solve overruns. Each result is tagged with the capture time of the frame from which it was computed, reaches the controller a median of 33.8 ms after that capture, and is held for up to one frame interval.
The cue uses the depth at which confidence falls below a threshold ,
with the number of rows, and takes its least-squares slope against normalized lateral position , negated so that a more confident left side gives a positive value (Fig. 2):
Because the map normalizes image intensity, a uniform change in scanner gain leaves it unchanged, and because (6) is a slope, a change in depth common to all columns leaves I unchanged. Map parameters [23] were selected offline on recorded scans: the depth attenuation , the edge-weight sharpness , the lateral edge penalty , and .
The cue used for control, , is passed through a second-order Butterworth low-pass filter with a 10 Hz cutoff, stepped at 1 kHz on the most recent frame value. Its group delay near DC, , is compensated in the delay reconstruction of Section IV.
The cue presupposes that most of the aperture is coupled. If contact is lost across a large fraction of the face, I remains well defined but no longer reports misalignment. This is a limitation of the cue rather than a condition the controller detects; the tilt limit of Section IV bounds how far a misreported cue can drive the probe.
IV. DUAL-RATE CONTROL
A. Fast Force Loop
The force loop regulates to a constant reference by commanding beam-axis velocity at 1 kHz (Fig. 3),
with saturated at 0.04 m/s and the derivative taken on the measurement. No integral term is used: the plant integrates the velocity command, so the loop already rejects constant force disturbances. The gains ( , ) were held fixed across every reported condition, and a safety stop ends a run if exceeds 30 N.
Locally, the contact obeys , where is the contact stiffness, the probe's rotation rate in the image plane, and the rotation-to-force gain defined below. With (7), the force error then evolves as
with the fast loop's disturbance-rejection rate and . is identified once per contact from the loading ramp, by a least-squares fit of against face depth along the beam axis (from forward kinematics, low-pass filtered with the same Butterworth filter as the joint torques) over the samples between and touchdown, when first comes within 10% of ; a fit with fewer than 100 samples or is rejected in favor of a

Fig. 3. Dual-rate control architecture. Dashed orange: once per image frame (30 Hz); solid blue: every 1 kHz tick; gray: physical and sensor paths. The cue is filtered and reconstructed to the present instant, the orientation command is scaled through the force-excursion constraint, and the force loop regulates the estimated contact force; is the constant 5 mm/s scan feed.
prior. The ramp precedes the scan feed, so surface elevation does not enter the fit. At the median per-contact stiffness of the comparison runs, (Section V), the loop gives and .
B. Orientation Channel
The orientation loop rotates the probe about the elevational axis of its face, changing its tilt in the image plane (measured from the touchdown orientation), so as to drive the cue to zero,
with cue gain rad/s, the sign of the cue sensitivity defined below, and saturated at . Net tilt from the touchdown orientation is limited to 0.55 rad (31.5°).
The law (9) acts directly on the cue and needs no conversion to angle. Controller P below uses two model quantities: the cue sensitivity , which converts elapsed rotation into an equivalent change in the cue, and the rotation-to-force gain at fixed face position, which converts a candidate rotation into a force excursion. is estimated online by recursive least squares with forgetting ( ), initialized at a prior set before the calibration of Section V, with its sign fixed from calibration and updates made only when the tilt has changed by more than over the preceding nine frames ( ). An underestimated under-compensates the delay, placing the reconstruction between none (as in B2) and full, and biasing the reconstruction comparison toward no difference. In the prediction, it slows the cue's decay and so overpredicts the excursion, as does the gain bound. is not estimated online: replaying the estimator on earlier phantom logs showed it is not identifiable from the controller's own rotations. It is instead held at a conservative bound , above every value measured in Section V.
C. Compared Controllers
All controllers share the force loop, tilt limit, and cue.
B1 applies no orientation correction, .
B2 applies (9) to the most recent filtered cue , with no delay reconstruction and no force constraint.
P, the proposed controller, adds two terms. The first is delay reconstruction. The cue was measured at capture time and the probe has rotated since, so the controller acts on
with read from a 1 kHz history buffer; the shift by compensates the filter's group delay. The second is a force-excursion constraint with fast-loop credit, which scales the commanded rate by . At each new frame, the controller simulates for a candidate , over a horizon ms, about one frame interval plus the capture-to-use delay (Section III),
from , where and is the component of caused by the rotation over the horizon. Because (8) is linear, is the sum of and the existing error, which decays at rate whether or not the probe rotates; the constraint therefore bounds only , and removing the existing error is left to the force loop. The second equation lets the commanded rate fall as the cue is nulled within the horizon. With the normalized simulated peak for a candidate , the applied scale is
held until the next frame, with and . The scalar minimization is solved by a bounded golden-section search over the feasible range of . The first term is the cost of withholding correction: zero at full correction and largest at none. The second is a logarithmic barrier on the excursion budget : negligible while the predicted excursion is small, so when the correction is harmless, and unbounded as the prediction approaches , so no minimizer can exceed it. The scale therefore varies continuously with the prediction rather than switching at the boundary; continuity matters because the prediction inherits the frame-to-frame variability of the cue.
The term is what distinguishes P from a one-step constraint. A one-step constraint assumes the excursion accumulates unopposed, over the horizon, whereas P credits the fast loop, which limits it to , so the correction is throttled less. The ablations change one setting each. (no barrier) computes but does not apply it, isolating the constraint from reconstruction. holds at its prior. is with (and ), the one-step (no-rejection) constraint. is B2 with fixed at 0.20 rad/s, near (14), testing whether a fixed cap reproduces the constraint.
D. Activity of the Constraint
At full rate, the constraint modifies the command appreciably only if the predicted excursion approaches or exceeds the budget,
The rotation limit is the worst case, and at this operating point also the actual one, since the rate saturates immediately after release. Condition (13) can be checked before experiments: any of a fast force loop, a slow rotation limit, or a wide budget makes the constraint inactive, in which case P reduces to B2 with reconstruction, while compliant contacts, reduced force gains, and lower-bandwidth force loops enlarge the regime in which it acts. We use rad/s and N, for which (13) holds with (predicted excursion 1.64 N at full rate and the median per-contact K), and report the realized scale s to confirm that the constraint acts. The realized scale reached its minimum at the first evaluation after release, (median over the 40 paired runs of P; range 0.265–0.276), and recovered to only after 2.6 s, once the cue had largely converged, so the constraint acts over most of cue convergence rather than only at engagement. is a budget on the excursion a rotation adds within one horizon, not a force-tracking band.
While the command is saturated, the constraint therefore acts as a ceiling on the rotation rate. Equating the predicted excursion to the budget gives the largest admissible rate,
which is 0.183 rad/s for P and at the median per-contact N/m (Section V), and rad/s for , independent of K.
Proposition 1: Under (8), if , , and on , then the rotation-induced error satisfies on . Proof: By (8) with , by (14).
Because each prediction starts from , sustained rotation at lets the rotation-induced error approach N under the gain bound. Minimizing (12) with at saturated demand places the rate below , at 0.163 and 0.108 rad/s; the logged medians over the first 0.5 s were 0.162 (P), 0.163 ( ), and 0.108 rad/s ( ). 's 0.20 rad/s cap lies 9% above P's and 23% above P's realized rate.
V. CHARACTERIZATION
All phantom experiments used a cylindrical gelatin phantom of radius 35 mm, prepared from water, gelatin, and cornstarch (a speckle agent) in an 80:10:1 weight ratio, with coupling gel at the interface. Scanner gain, dynamic range, time-gain compensation, speckle reduction, and adaptive gain were fixed for all runs, since the cue is comparable only within a fixed display configuration.
A. Contact Stiffness
The three loading ramps of the tilt sweeps below gave (median; range 1279–1395, ), which set the prior. Over the loading ramps of the comparison runs, the per-contact fit, which the controller used, gave (median; range 1068–1199), the value at which the rates of Section IV are evaluated; no run in the reported comparisons fell back to the prior.
B. Tilt Sweep
With the orientation loop disabled, the probe was stepped through face-pivot angles out to and back while holding 15 N, dwelling 3 s at each angle, in three sweeps (93 dwells). The force loop returns to at every dwell (mean error below 0.01 N), so is not observable as a force change; it appears as the depth change needed to maintain 15 N. With the face depth along the touchdown beam axis, removes the force explained by depth, leaving the tilt-dependent component up to a constant, and . No parametric angular-contact model is assumed: is estimated as the local slope of within fixed bands, after removing a slow drift (gel relaxation, N per 127 s sweep) that the ascending and descending passes separate.
Rotation-to-force gain. Within the swept range kept one sign: the banded slopes were all positive, between 11.5 and 26.3 N/rad (largest 95% bound 33.8 N/rad), a factor of 1.5 below (1.2 for the largest bound). The bound is thus verified only within ; at larger tilts it is assumed.
Cue sensitivity. Over the same dwells, the cue varied with tilt at (95% CI , ), 1.8 times the prior used by the controllers. Under P the online estimate moved toward it from -0.050 during roughly the first half of each run and then held near a median of (range -0.083 to -0.064 across runs), short of the calibrated value.
VI. CONTROLLER EXPERIMENTAL VALIDATION
A. Protocol
The controllers of Section IV were compared on the phantom of Section V, traversed along its axis for 100 mm at 5 mm/s with N and no lateral correction. The orientation demand arises at engagement: every run starts from the same pose, from which the cue reports a
TABLE I PAIRED DIFFERENCES BETWEEN CONDITIONS (A - B), MEAN AND PAIRED-t 95% CI, PAIRS EACH.
| A - B | (N) | (N) | (s) |
|---|---|---|---|
| B1 - B2 | -1.95*[-2.08, -1.82] | -2.07*[-2.19, -1.95] | +6.66*[-5.91, +7.40] |
| B2 - P | +0.40*[+0.13, +0.67] | +0.40*[+0.30, +0.51] | -0.94*[-1.04, -0.84] |
| - P | +0.49*[+0.29, +0.69] | +0.54*[+0.33, +0.75] | -0.93*[-1.00, -0.86] |
| B2 - | +0.11[-0.10, +0.32] | +0.07[-0.11, +0.25] | +0.01[-0.03, +0.05] |
| - P | +0.03[-0.20, +0.25] | +0.09[-0.08, +0.26] | -0.32*[-0.46, -0.19] |
| - | -0.96*[-1.20, -0.72] | +0.03[-0.16, +0.22] | +0.59*[+0.43, +0.76] |
| - P | -0.17*[-0.33, -0.01] | +0.05[-0.05, +0.16] | -0.01[-0.04, +0.02] |
*95% CI excludes zero.
misalignment, so the orientation loop rotates the probe as soon as it is released. By the time the cue reached 20% of its initial value, the probe had rotated a median of under P (20.0–26.9° across the correcting conditions), toward the orientation of highest confidence, which need not coincide with the geometric normal. In every condition, the orientation loop is held off for 1 s after touchdown, so that the force loop settles first and the touchdown transient is not confounded with rotation onset, and all outcomes are measured from that release. All parameters ( rad/s, N, N/rad, the prior S, force gains, cue parameters, and scanner configuration) were fixed before collection.
The seven comparisons of Table I were collected as seven blocks of 20 runs in interleaved ABBA order; each run is one experimental unit. Each run is paired with its time-adjacent partner of the other condition in the same comparison, giving 10 pairs per comparison, balanced in order.
The outcomes are the RMS force error over the first second after release, , which captures the engagement transient, together with the peak error over the same second; the RMS force error from release to the end of the traverse, E; and the cue-convergence time , the time from release until the magnitude of the filtered cue first falls to half its initial value (its mean over the first 30 ms). The residual cue, the median of over the last 5 s, shows whether low cue asymmetry is maintained. Table I reports paired differences. Confidence intervals quantify effect size and precision; no correction for multiple comparisons was applied.
B. Results
Cost of correction. B1 never rotates. Against B2 it lowers by 1.95 N, the peak error in the first second by 5.98 N, and E by 2.07 N, but it does not actively reduce the cue (Fig. 4b): its cue reached half its initial value only after a median of 7.4 s, without any correction, as the orientation of highest confidence shifted along the traverse, and ended at 0.009 against 0.002 for every correcting controller. Orientation correction is not free: rotating the probe perturbs contact force through . The remaining rows measure how that cost depends on the correction rate.
Rate constraint. Under B2 the probe rotates at the full rate from release. Relative to B2, P lowers by 0.40 N, the peak error by 2.31 N, and E by 0.40 N, and its cue converges 0.94 s later (median 1.62 against 0.69 s; Fig. 4b). Removing the constraint while keeping reconstruction and online estimation ( ) raises by 0.49 N, the peak by 2.74 N, and E by 0.54 N, and shortens by 0.93 s. and B2 do not differ resolvably on any outcome, consistent with the constraint accounting for the differences in both and E and with reconstruction alone having no resolvable effect. B2 and approached or reached the 31.5° tilt limit (median peaks 29.7° and 31.6°), whereas the rate-limited controllers stayed below 28°.
Constraint versus a fixed cap. B2 , whose cap lies within 9% of P's (14), does not differ resolvably from P in (+0.03 N) or E (+0.09 N), and its cue converges 0.32 s sooner (median 1.32 against 1.66 s). Both limits act on the same saturated demand. P's realized rate (0.162 rad/s, Section IV) lies 19% below the fixed cap, comparable to the 20% difference in cue-convergence time, and the lower rate bought no resolvable reduction in force error; at this operating point the fixed cap is the better choice.
Fast-loop credit and adaptation. Removing the credit ( against ) lowers the ceiling from 0.183 to 0.125 rad/s (realized rate 0.163 to 0.108 rad/s). It lowers by 0.96 N and the peak error by 2.70 N, does not change E resolvably, and lengthens by 0.59 s (median 2.12 against 1.61 s). Holding S at its prior ( ) lowers marginally by 0.17 N and changes no other outcome resolvably.
Ordering. Across blocks, median follows the realized rotation rate: 0.65–0.69 s for B2 and (0.6 rad/s), 1.32 s for (0.20), 1.52–1.66 s for P and (0.16, the rate the barrier predicts below ), and 2.12 s for (0.11). Within every block that pairs clearly different rate limits, the lower limit had the lower ; this ordering across blocks is not itself a paired comparison.
C. Volunteer Feasibility
A single-subject popliteal scan tests whether the same controllers, unchanged, operate on living tissue, which adds spatially varying stiffness and curvature and subject motion (arterial pulsation, muscle activity, and limb drift) not represented by the phantom. It is reported as feasibility evidence and is not pooled with the phantom statistics.
P and B2 were each run six times in interleaved order and paired as on the phantom. Relative to P, B2 raised by 1.44 N (95% CI [0.80, 2.09]), the peak error in the first second by 5.91 N ([3.32, 8.50]), and E by 1.05 N ([0.82, 1.29]). Median E was 2.02 N under P and 3.09 N under B2. On tissue the cue was dominated by noise: its magnitude at release (median 0.013–0.020, against 0.08 on the phantom) was comparable to its residual over the traverse (0.017–0.019), so convergence times on tissue do not measure alignment and are not reported. Before the reported runs, the probe placement was adjusted over several setup attempts, some of which ended at the force-limit safety stop, often when the rotation halted abruptly at the tilt limit; these attempts are not analyzed. The human subject gave informed

(a)
(b)
Fig. 4. Engagement and cue convergence after release (t = 0) on the phantom. (a) Running RMS of force error (100 ms centered window); shading marks the per-horizon budget N. (b) Filtered cue magnitude over its mean in the first 30 ms, ; dotted lines mark 50% ( ) and 20%. Line styles as in (a). Curves are medians over the 10 paired runs of each condition in one block (B2: B2/P; P and B2 : B2 /P; P : P /P ; B1: B1/B2), with interquartile bands. Offsets between curves from different blocks are not paired comparisons; the paired comparisons are in Table I. P and P (omitted) overlap B2 and P, respectively.
consent under a protocol approved by the University of Alabama Institutional Review Board (protocol ID 24-01-7230). These data do not establish clinical accuracy, anatomical generalization, or diagnostic image quality.
VII. Discussion
The contribution is not the combination of image and force feedback, which is established [8], [13], [14], but a principled way to set how fast an image-guided correction may rotate the probe under force control. The coupling model gives the ceiling of (14) in closed form, and the experiments support it in two ways. Cue-convergence time followed the realized rate, and in every block pairing clearly different rate limits, the lower one had the lower engagement error. A fixed cap of 0.20 rad/s, within 9% of the predicted ceiling , matched the online predictive constraint in force error, neither paired difference resolvable (upper 95% bounds 0.25 and 0.26 N), while converging 0.32 s sooner. The ceiling therefore carries the force benefit of the prediction and can be imposed directly: compute from the contact stiffness, the force-loop gains, the gain bound, the budget, and the horizon, and cap the rotation rate at it. Delay reconstruction had no resolvable effect, although S ran at 57–82% of its measured value; it also had little to correct: under saturated demand the command ignores the cue's value, and in the proportional regime the orientation loop crosses over at rad/s, where the 55–90 ms capture-to-use delay (Section III, with ) costs only 10–20° of phase, not counting the scanner's unresolved internal delay (Section II). An online constraint can improve on the fixed cap where the demand falls below the ceiling or where stiffness or gain changes within a scan, which a single phantom did not test.
The rotation-to-force gain depends on surface geometry and contact state. On the cylindrical phantom, it kept one sign and stayed at or below 26.3 N/rad over the sweep (Section V), a narrower range than the rotations of Section VI, and it was not identifiable online from the controller's own rotations.
The prediction therefore uses a bound on the gain rather than a model of it. The gain bound N/rad exceeds every band gain measured on the phantom (largest 95% bound 33.8 N/rad), so the predicted excursion, and with it the admissible rate , errs toward throttling. Where the true gain is smaller, the bound costs cue-convergence speed. Because comparison-run stiffness varied over only 1068–1199 N/m, the predicted dependence of on K was validated only locally around this operating region.
Several limitations qualify these results. The contact-force estimate has no independent reference during operation and inherits the accuracy bound of Section II, so all force results are stated in estimate space rather than as calibrated force, and they include the effect of the arm dynamics that the quasi-static estimator omits (Section II). The budget N, comparable to that accuracy (0.41 N RMS), constrains changes in the estimate rather than in the true force, whose excursion is not independently verified. The confidence map's principal weakness is that a uniformly anechoic region presents no gradients to the random walk and may be assigned high confidence, so the cue reports asymmetry reliably but is not a contact detector. The volunteer scan exposed a second, related weakness. Over the curved popliteal fossa, one edge of the aperture lost acoustic coupling for sustained periods, and the loop did not null the resulting misalignment. The cue assumes that confidence differs between the two sides of the image mainly because of coupling; in heterogeneous tissue, attenuation and structure also lower and spread confidence on the coupled side, which likely diluted the asymmetry the decoupled edge should have produced. The cue is therefore insufficient for alignment on heterogeneous tissue. Surfaces on which the true gain approaches , such as stiffer tissue or sharper curvature, would need a different bound and were not tested.
On tissue the cue was noise-dominated, its magnitude at release comparable to its residual, yet P still lowered E by 1.05 N relative to B2, which makes the cue rather than the architecture the binding limitation on living tissue. The cue is low-pass filtered and the filter's group delay is compensated in (10), but residual cue noise on tissue still converts directly into commanded rotation, and the tilt limit bounds the consequence without addressing the cause. A confidence or cue measure that adapts to heterogeneous tissue rather than one tuned on homogeneous gelatin, or an explicit detector for edge decoupling, is the natural next step and is not attempted here. Finally, characterization is confined to homogeneous gelatin, and the volunteer scan is a single-subject feasibility test that establishes neither clinical safety nor anatomical generalization.
VIII. Conclusion
We analyzed the coupling between a high-rate force loop using joint-torque-based estimation and a slower, delayed image-guided orientation loop in robotic ultrasound. Under saturated demand, the proposed constraint, which credits the fast loop's rejection over the prediction horizon, reduces to a closed-form rotation-rate ceiling computable before each scan from identified system parameters.
Across seven paired comparisons on a curved phantom, orientation correction carried a measurable force cost, and the ceiling set the trade-off between cue-convergence time and force error. A fixed cap near the predicted rate matched the online constraint in force error and converged faster, and delay reconstruction had no resolvable effect. For approximately stationary contact properties, the closed-form cap is therefore preferable to the online predictor because it provides comparable force regulation with lower implementation complexity. On a single-subject popliteal scan, the rate-limited controller lowered the RMS force error E by 1.05 N relative to the unconstrained baseline, although the image cue was noise-dominated on tissue. Future work includes developing an image cue robust to heterogeneous tissue and testing on more subjects and anatomies.
ACKNOWLEDGMENT
The authors acknowledge financial support from The University of Alabama faculty startup fund and the Alabama Transportation Institute. This work was also partly supported by the U.S. Department of Transportation University Transportation Centers Program through CR2C2, Grant No. 69A3552348304. Claude [24] was used under the authors' direction to assist with experiment planning, data analysis, code development, figure preparation, and manuscript drafting and editing, and ChatGPT [25] to assist with manuscript review and editing. The authors built the experimental system and conducted and supervised all experiments; AI-assisted code was validated by testing on that system and by checking its outputs, and the authors take full responsibility for the content of this paper.
REFERENCES
[1] K. Li, Y. Xu, and M. Q.-H. Meng, “An overview of systems and techniques for autonomous robotic ultrasound acquisitions,” IEEE Transactions on Medical Robotics and Bionics, vol. 3, no. 2, pp. 510–524, 2021.
[2] S. E. Salcudean, H. Moradi, D. G. Black, and N. Navab, “Robot-assisted medical imaging: A review,” Proceedings of the IEEE, vol. 110, no. 7, pp. 951–967, 2022.
[3] Z. Jiang, S. E. Salcudean, and N. Navab, “Robotic ultrasound imaging: State-of-the-art and future perspectives,” Medical Image Analysis, vol. 89, p. 102878, 2023.
[4] T. Han, G. Ning, H. Liang, Z. Li, Z. Jiang, F. Chen, Y. Kang, J. Luo, and H. Liao, “Toward clinical applications of intelligent robotic ultrasound systems,” IEEE Reviews in Biomedical Engineering, vol. 19, pp. 227–247, 2026.
[5] K. Munir, A. F. Al-Battal, A. Alsheghri, H. Becher, M. Noga, and K. Punithakumar, “A survey of autonomous robotic ultrasound scanning systems,” IEEE Access, vol. 13, pp. 103 178–103 197, 2025.
[6] P. Abolmaesumi, S. E. Salcudean, W.-H. Zhu, M. R. Sirouspour, and S. P. DiMaio, “Image-guided control of a robot for medical ultrasound,” IEEE Transactions on Robotics and Automation, vol. 18, no. 1, pp. 11–23, 2002.
[7] P. Chatelain, A. Krupa, and N. Navab, “Optimization of ultrasound image quality via visual servoing,” in Proceedings of the IEEE International Conference on Robotics and Automation, 2015, pp. 5997–6002.
[8] Z. Jiang, M. Grimm, M. Zhou, J. Esteban, W. Simson, G. Zahnd, and N. Navab, “Automatic normal positioning of robotic ultrasound probe based only on confidence map optimization and force measurement,” IEEE Robotics and Automation Letters, vol. 5, no. 2, pp. 1342–1349, 2020.
[9] Z. Jiang, M. Grimm, M. Zhou, Y. Hu, J. Esteban, and N. Navab, “Automatic force-based probe positioning for precise robotic ultrasound acquisition,” IEEE Transactions on Industrial Electronics, vol. 68, no. 11, pp. 11200–11211, 2021.
[10] X. Ma, W.-Y. Kuo, K. Yang, A. Rahaman, and H. K. Zhang, “A-SEE: Active-sensing end-effector enabled probe self-normal-positioning for robotic ultrasound imaging applications,” IEEE Robotics and Automation Letters, vol. 7, no. 4, pp. 12475–12482, 2022.
[11] A. Piedra, R. B. Jeffrey, and O. Khatib, “Explicit force-based control strategies for compliant robot-assisted ultrasound imaging,” The International Journal of Robotics Research, 2026, early access, doi: 10.1177/02783649261441635.
[12] Z. Jiang, Z. Li, M. Grimm, M. Zhou, M. Esposito, W. Wein, W. Stechele, T. Wendler, and N. Navab, “Autonomous robotic screening of tubular structures based only on real-time ultrasound imaging feedback,” IEEE Transactions on Industrial Electronics, vol. 69, no. 7, pp. 7064–7075, 2022.
[13] M. Akbari, J. Carriere, T. Meyer, R. Sloboda, S. Husain, N. Usmani, and M. Tavakoli, “Robotic ultrasound scanning with real-time image-based force adjustment: Quick response for enabling physical distancing during the COVID-19 pandemic,” Frontiers in Robotics and AI, vol. 8, p. 645424, 2021.
[14] E. Zakeri, A. Spilkin, H. Elmekki, A. Zanuttini, L. Kadem, J. Bentahar, W.-F. Xie, and P. Pibarot, “AI-powered robust interaction force control of a cardiac ultrasound robotic system,” IEEE Transactions on Industrial Electronics, vol. 72, no. 4, pp. 3972–3983, 2025.
[15] H. Hase, M. F. Azampour, M. Tirindelli, M. Paschali, W. Simson, E. Fatemizadeh, and N. Navab, “Ultrasound-guided robotic navigation with deep reinforcement learning,” in Proceedings of the IEEE/RSJ International Conference on Intelligent Robots and Systems, 2020, pp. 5534–5541.
[16] G. Ning, X. Zhang, and H. Liao, “Autonomic robotic ultrasound imaging system based on reinforcement learning,” IEEE Transactions on Biomedical Engineering, vol. 68, no. 9, pp. 2787–2797, 2021.
[17] H. Jiang et al., “Towards expert-level autonomous carotid ultrasonography with large-scale learning-based robotic system,” Nature Communications, vol. 16, p. 7893, 2025.
[18] A. Duan, C. Yang, J. Zhao, S. Huo, P. Zhou, W. Ma, Y. Zheng, and D. Navarro-Alarcon, “Safe learning by constraint-aware policy optimization for robotic ultrasound imaging,” IEEE Transactions on Automation Science and Engineering, vol. 22, pp. 2349–2360, 2025.
[19] S. Li, Y. Pi, W. Li, G. Carbone, and K. Wu, “Trajectory-aware feedforward admittance control for contact force tracking with time-varying reference trajectory in a teleoperated robotic ultrasound system,” Control Engineering Practice, vol. 168, p. 106674, 2026.
[20] Y. Qian, J. Luo, Y. Zhang, K. Long, Y. Yuan, M. Q.-H. Meng, and L. Liu, “Symmetry-optimized image guidance and biomechanics-driven force measurement for robotic ultrasound: A control framework enhancing sonographic quality,” IEEE Transactions on Instrumentation and Measurement, vol. 75, p. 7504018, 2026.
[21] R. Featherstone, Rigid Body Dynamics Algorithms. New York: Springer, 2008.
[22] M. Brokate and J. Sprekels, Hysteresis and Phase Transitions. New York: Springer, 1996.
[23] A. Karamalis, W. Wein, T. Klein, and N. Navab, “Ultrasound confidence maps using random walks,” Medical Image Analysis, vol. 16, no. 6, pp. 1101–1112, 2012.
[24] Anthropic, “Claude,” Large language model, https://claude.ai, 2026, accessed Oct. 2026.
[25] OpenAI, “ChatGPT,” Large language model, https://chatgpt.com, 2026, accessed Oct. 2026.