arXiv:2610.06010cs.ROOct 2026

Manuscript Offprint

arXiv:2610.06010

Recon2Servo: Robotic Ultrasound Visual Servoing via Learned Image-to-Motion Inference

Yameng Zhang, Pei Liu, Dianye Huang, Yizhao Qian, Zhongyu Chen, Xiangyu Chu, K. W. Samuel Au, Zhongliang Jiang

Yameng Zhang 1,2,3^{1,2,3}, Pei Liu 1,4^{1,4}, Dianye Huang 1^{1}, Yizhao Qian 5^{5}, Zhongyu Chen 6^{6}, Xiangyu Chu 2,3^{2,3}, K. W. Samuel Au 2,3^{2,3}, and Zhongliang Jiang 1,∗^{1,*}

Abstract—Ultrasound visual servoing is essential for autonomous robotic ultrasound, yet 6-DoF probe control from 2D B-mode images remains challenging due to limited and ambiguous out-of-plane motion cues. Existing methods typically rely on anatomical priors or handcrafted visual features, limiting their generalizability across imaging targets. Inspired by trackerless 3D ultrasound reconstruction, we propose Recon2Servo, a visual servoing framework that learns image-to-motion inference directly from B-mode images for 6-DoF probe control. A DINOv3 encoder with low-rank adaptation and a bidirectional relation module estimate the relative probe pose between current and target images to guide iterative closed-loop target-view alignment. The framework combines supervised relative-pose learning, reconstruction-guided closed-loop adaptation, and bounded residual pose correction to improve motion inference during servoing. Evaluations on a public dataset and an in-house dataset collected from 12 healthy volunteers using different ultrasound systems demonstrate its effectiveness in reconstructed-volume servoing. Additional real-robot demonstrations of target-view alignment and dynamic tracking on a human forearm are provided in the supplementary video: https://youtu.be/qwsOdI-GMYk.

I. INTRODUCTION

Ultrasound visual servoing is a critical technology for robotic ultrasound, as it enables the robot to directly exploit real-time imaging feedback to regulate probe motion according to the observed anatomy, thereby supporting autonomous probe navigation, image stabilization, and target tracking [1]. However, conventional medical ultrasound typically provides cross-sectional 2D images. While in-plane motion can be relatively easily estimated from abundant visual features, out-of-plane probe motion is long-standing challenging to infer because it causes anatomical structures to disappear from or newly enter the imaging plane, providing limited and often ambiguous motion cues.

In a pioneering study, Abolmaesumi et al. proposed a shared-autonomy strategy combining human expertise with ultrasound visual servoing [2]. The in-plane 3-DoF probe motion was automatically controlled to center the carotid artery in cross-sectional ultrasound images, while the remaining three out-of-plane DoFs were teleoperated by a human expert. Recently, Qian et al. employed optical flow as a feature for high-frequency 3-DoF translational tracking [3]. Extending ultrasound visual servoing to fully control the 6-DoF probe motion, however, requires servoing features that encode both in-plane and out-of-plane motion. In 2006, Bachta et al. developed a servoing method to drive the probe from an arbitrary initial pose toward a desired ultrasound image [4]. The features were derived from object boundaries extracted from ultrasound images, with the target contour recovered using polynomial regression tailored to simplified egg-shaped objects.

To enable generic visual servoing without prior knowledge of object geometry, Krupa et al. exploited ultrasound speckle to estimate both in-plane and out-of-plane probe motion [5], [6]. Speckle patterns exhibit spatial coherence across consecutive ultrasound images, providing intrinsic visual cues for estimating probe motion. However, speckle also degrades image quality and is therefore commonly suppressed by despeckling and image-enhancement techniques in modern ultrasound systems [7], limiting its reliability and generalizability as a visual servoing feature. Beyond speckle-based approaches, alternative image representations have been explored to construct features for 6-DoF visual servoing. Mebarki et al. employed image moments to characterize the spatial distribution of image content [8], [9], while Nadeau et al. directly exploited image intensities as servoing features [10]. Moving toward less handcrafted feature representations, Duflot et al. first introduced subsampled shearlet coefficients as visual features for the control law [11]. By providing a redundant yet noise-robust representation of ultrasound image content in a selected frequency domain, the shearlet coefficients enabled accurate, repeatable, and robust visual servoing, as demonstrated in both simulation and phantom experiments.

With the rapid advances in deep learning, recent studies have increasingly explored reinforcement and imitation learning for standard-plane navigation. Li et al. formulated 6-DoF standard-plane navigation as a reinforcement learning problem with a handcrafted reward function [12]. To reduce the reliance on manually designed rewards, Jiang et al. employed inverse reinforcement learning to infer image-based rewards from a small number of expert demonstrations [13]. H. Jiang et al. further adopted imitation learning to directly learn probe navigation from demonstrations, demonstrating its effectiveness in standardized carotid ultrasound scanning workflows [14]. However, these approaches are typically trained for predefined target planes or scanning tasks and therefore cannot readily generalize to arbitrarily specified

*Corresponding author: Zhongliang Jiang (email: zlijiang@hku.hk).

1^{1}Department of Mechanical Engineering, The University of Hong Kong, Hong Kong SAR. 2^{2}Multi-scale Medical Robotics Center, AIR@InnoHK, Hong Kong SAR. 3^{3}Department of Mechanical and Automation Engineering, The Chinese University of Hong Kong, Hong Kong SAR. 4^{4}School of Computation, Information and Technology, Technical University of Munich, Germany. 5^{5}Department of Electronic Engineering, The Chinese University of Hong Kong, Hong Kong SAR. 6^{6}School of Optometry, Hong Kong Polytechnic University, Hong Kong SAR.

Figure

Fig. 1. Overview of the proposed ultrasound servoing framework. Tracked B-mode images are reconstructed into a 3-D volume, which serves as an interactive simulation environment for training and adapting the probe control policy. The resulting policy is deployed using real-time ultrasound image feedback to iteratively guide the probe toward a target view.

imaging planes, due to the lack of a reliable mapping between ultrasound image variations and the corresponding spatial probe motion.

To address this challenge, we draw inspiration from trackerless freehand 3D ultrasound reconstruction, which has attracted increasing attention in recent years. At its core, trackerless reconstruction aims to infer the spatial motion of the ultrasound probe directly from image sequences, as demonstrated by early inter-frame pose regression methods [15]. Recent studies have further improved motion estimation by modeling long-term sequence dependencies [16], incorporating physics-guided speckle correlation [17], learning fine-grained spatiotemporal representations with inertial alignment [18], and separately encoding local motion and global anatomical context, as in DualTrack [19]. Collectively, these advances demonstrate that spatial probe motion can be inferred from ultrasound image sequences by jointly exploiting local image variations and broader anatomical context. Inspired by these advances, we explore whether the image-to-motion inference capability developed for trackerless reconstruction can be leveraged to establish a generalizable spatial representation for ultrasound visual servoing.

In this study, we propose Recon2Servo, a novel ultrasound visual servoing framework that directly learns 6-DoF probe motion from B-mode ultrasound images. To the best of our knowledge, Recon2Servo is the first approach to achieve 6-DoF ultrasound visual servoing through learned image-to-motion inference, without relying on sophisticated handcrafted visual features, whose design is particularly challenging in ultrasound due to substantial inter-operator variability and inherent imaging noise. Specifically, a DI-NOv3 encoder [20], adapted using low-rank adaptation (LoRA) [21], extracts multi-level visual features that are fed into a bidirectional relation module with dedicated translation and rotation motion tokens to infer the relative 6-DoF pose between the current and target images. The predicted relative pose is directly converted into probe motion through a fixed-gain controller. To improve robustness to states encountered during closed-loop servoing, we further introduce a dataset aggregation (DAgger)-style adaptation strategy [22], which updates the pose estimator using labeled states visited during tracked reconstruction trajectories. Finally, a bounded residual head leverages frozen relational features and the mean pose estimate to further refine the motion prediction. To evaluate the proposed method, we assess relative-pose estimation accuracy and closed-loop target-view convergence using the public TUS-REC2024 dataset [23] and an in-house ultrasound dataset collected from 12 healthy volunteers. These evaluations examine the effectiveness of the learned image-to-motion representation and its adaptation across ultrasound acquisition domains.

References (23)