arXiv:2610.02668cs.ROOct 2026

Manuscript Offprint

arXiv:2610.02668

A Passive AI System for Verifying Physical State on Automated Liquid Handlers

Junqiong Joanne Qiu, Zeckria Kamrany, Ananya Anand, Emma Vidal, Yunjia Johanna Bai, Nathan S Chen, Sean Son, Varada Abhyankar, Taryn Jakub, Eleazar Eskin

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Abstract

Automated liquid handlers execute digital protocols, but operators must assemble the deck and confirm that the physical setup matches the intended experiment. We present the Labware Setup Checker, a passive vision system that verifies deck preparation on an Opentrons Flex without modifying its hardware or firmware. A consumer webcam mounted outside the robot supplies images to a browser application. The checker segments the deck into slots, corrects slot assignments using the known deck geometry, classifies the labware, and compares the observed setup with a reference protocol library to report missing, misplaced, or unexpected labware. It operates independently of the robot controller and does not interrupt or gate a run. During SwabSeq Respiratory Viral Panel (RVP) pre-PCR deck preparation, the checker tracked a slot from empty to a base and then to the completed plate-and-base assembly, identifying the completed assembly in all 29 frames before robot motion. Across these frames, 95.1% of the 348 labware classifications for slots requiring exact matches agreed with the protocol reference. In 480 frames with partially occluded reservoirs, all 6,720 slot assignments were correct. Across three camera positions and two lighting conditions, slot assignment was 100% correct, and 97.73% of labware classifications matched the protocol reference. Median verification time was 2.10 s on a laptop using only its CPU and 4.24 s from frame capture to receipt of results during AWS deployment in the laboratory. These results show that an external camera and independent software can provide verification of deck setup against protocol requirements within seconds, allowing operators to review the physical setup before starting a run.

Keywords: laboratory automation, computer vision, liquid handling, Opentrons, quality control, protocol verification

1 Introduction

Laboratories are increasingly automating routine workflows. In a 2025 survey, 49% of respondents reported using robotics or automation in their company's laboratories [1]. Automated liquid handlers are a key tool in laboratory automation, supporting molecular biology workflows such as sequencing library preparation, PCR setup, and high-throughput processing [2]. They enable workflows to be run reproducibly with minimal manual labor. These systems are offered by manufacturers such as Opentrons, Tecan, and Hamilton. Quality-control studies of laboratory automation show that liquid handlers require routine checks to stay accurate and reliable, since components can fail over time [3, 4]. Setup errors, however, can occur even earlier: a plate placed in the wrong slot, the wrong tip rack loaded, a reservoir missing, or labware shifted or occluded on the deck. These errors can cause collisions, damage the robot, or lead to incorrect liquid transfers that compromise the experiment. This physical starting state requires manual assembly: labware, reagents, plates, and tip racks must be placed on the deck in the correct locations before the protocol begins. Correct deck setup relies on user confirmation, and there is no system in place to confirm that the setup matches the robot's expectations, making the initial manual deck setup step prone to errors. Therefore, a monitoring layer is needed to answer three questions: what experiment the operator is running, whether the deck is prepared correctly for it, and whether the robot is executing the protocol correctly.

To implement a monitoring layer, a visual system is required. Currently, camera-based deck checking is available on the Opentrons Flex, which carries a built-in 2 MP camera that records images and video [5]. The camera sends images to an operator or to storage, but does not report whether the observed labware is correct for a given protocol. Tecan Fluent workstations run DeckCheck, which photographs the worktable after loading and compares the live layout against a stored reference [6]. However, DeckCheck only compares pixels and not labware identity, leading to the manual listing differences it cannot flag: different tip types, missing tubes on partially loaded runners, plates rotated by 180 degrees, and well shape [6]. Research systems have used computer vision to monitor laboratory processes and detect anomalies during automated liquid transfer [7, 8], and have supported colorimetric titration.

A PREPRINT

A. Checker interface

Figure

B. Incorrect deck setup

Figure

C. Returned verification result

Figure

Figure 1. Web interface and protocol-aware deck verification. A. The checker displays the live camera feed with predicted slot and labware labels, status updates, detection logs, and the predicted protocol. B. An incorrect pre-PCR deck setup in which B3 is empty and the expected 50 μ\muL tip rack is placed at C3. C. The returned verification result. Green boxes indicate that the observed labware matches the protocol or that an expected-empty slot is empty. The red box at B3 indicates that the expected tip rack is missing, and the orange box at C3 indicates that the tip rack was detected in the wrong slot.

[9], but these remain isolated capabilities rather than a general check against a protocol’s requirements.

Existing options do not provide a modular and accessible verification layer that a laboratory can add to the liquid handler it already owns. DeckCheck is integrated into a high-end workstation and is unavailable for lower-cost platforms. Accessible self-driving laboratories close the loop between planning, execution, measurement, and decision-making [10–12][10–12], but also require tightly integrated instruments, data infrastructure, and expertise that academic life-science laboratories often lack [13][13]. As a result, a laboratory running an affordable liquid handler has no straightforward way to check its deck against the intended protocol.

Such a system must meet four constraints: it must be inexpensive, leave robot hardware and firmware untouched, run on an available computer, and add no work to the operator's routine. Computer vision makes this checking system practical: a camera is an inexpensive non-contact sensor, and deep learning now localizes, segments, and classifies objects reliably in visually cluttered scenes [14,15][14, 15]. Applied to a deck image, these methods can find each slot, separate labware from its background, identify the labware type, and map it to the robot's canonical deck coordinates, turning a photograph into a structured deck state that can be compared with a protocol.

We present the Labware Setup Checker, a passive vision system for the Opentrons Flex that utilizes a low-cost camera and lightweight models to confirm an observed deck setup against an intended protocol within seconds. A consumer webcam outside the robot observes the deck. The checker segments the image into slots, classifies masked and cropped labware images, corrects the assignments against canonical slot geometry, and compares the result with a reference protocol library, marking each slot as correct, wrong, misplaced, or unexpected. The checker operates passively and noninvasively, requiring only a simple session setup and checking the deck automatically without interrupting robot operation or adding manual steps to the established procedure. This work contributes an external verification architecture for the Opentrons Flex that requires no hardware or firmware modification, protocol-aware deck perception at seconds-scale latency in a deployed setting, and an evaluation system compatible with routine laboratory workflows.