Manuscript Offprint

arXiv:2608.23432

Image-Conditioned Diffusion Models for Quality Assurance of Organ-at-Risk Segmentations in Radiotherapy

Clea Dronne, Catharine H Clark, Xavier Loizeau, Elizabeth Miles, Peter Hoskin, Jamie R McClelland

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

C Dronne 1,3^{1,3}, C H Clark 2,3,4^{2,3,4}, X Loizeau 3^{3}, E Miles 5^{5}, P Hoskin 5,6^{5,6}, and J R McClelland 1^{1}

1^{1} UCL Hawkes Institute, Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom

2^{2} Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom

3^{3} National Physical Laboratory, Teddington, UK

4^{4} Department of Radiotherapy Physics, University College London Hospital, London, UK

5^{5} National Radiotherapy Trials Quality Assurance Group (RTTQA), Mount Vernon Hospital, Northwood, UK

6^{6} Division of Cancer Sciences, University of Manchester, Manchester, UK

Abstract. Accurate organ-at-risk segmentation is essential for radiotherapy planning, but reviewing segmentations is time-consuming and subjective. We investigate normative modelling for segmentation error detection in head-and-neck CT, comparing a VAE framework with an image-conditioned segmentation diffusion model. Models were evaluated on RADCURE brainstem and spinal cord segmentations using simulated boundary and width perturbations. Error detection was assessed using the Dice similarity coefficient and the Distance to Agreement (DTA) between the input and reconstructed segmentations. While both models detected some simulated errors, regional DTA showed that the diffusion model localised subtle boundary errors more consistently. These results support image-conditioned diffusion reconstruction as a promising framework for localised, anatomy-aware segmentation QA. Code is available here.

Keywords: error detection · diffusion models · radiotherapy.

1 Introduction

Accurate delineation of organs at risk (OARs) is essential for radiotherapy treatment planning and clinical trial quality assurance (QA). Segmentation errors can reduce tumour dose coverage and increase toxicity to healthy tissues. In current clinical practice, OAR segmentations may be produced manually, generated by automated models, or created through hybrid AI-assisted workflows, but all require review before clinical use. In clinical trials, where consistency is critical, this review remains a time-consuming and subjective manual process [14][14].

Dronne et al.

Automated segmentation QA aims to support expert review by identifying cases or regions that may require closer inspection. Unlike uncertainty quantification methods, which are typically tied to a specific segmentation model, post-hoc QA methods should be model-agnostic and applicable to manual, AI-generated, or edited segmentations [2].

Prior work shows that anatomical image context is essential, as shape-only methods are insufficient for assessing whether a segmentation is appropriate for a specific patient anatomy [5,9]. Heuristic or summary metrics likewise fail to capture full spatial information and may miss clinically relevant errors [3,11].

Normative modelling provides one approach to segmentation QA by learning the distribution of clinically acceptable image-segmentation pairs. Given a CT image and a segmentation under review, the model reconstructs a segmentation consistent with the learned distribution, and discrepancies between the reconstruction and the submission can serve as a QA signal. Previous studies have demonstrated the feasibility of this framework using Variational Autoencoders (VAEs) [7,12,15].

Denoising Diffusion Probabilistic Models (DDPMs) [8] offer an alternative generative framework that has achieved strong performance in image reconstruction and anomaly detection [1,18]. Their ability to generate sharp, spatially faithful reconstructions may be particularly valuable for segmentation QA, where errors often manifest as subtle local boundary deviations.

In this work, we compare a previously established VAE segmentation QA approach with a new image-conditioned segmentation diffusion model. We evaluate whether using a diffusion model improves the detection and localisation of simulated OAR segmentation errors in a 3D patch-based framework.

References (18)