Ex vivo breach detection using electrical conductivity during robotic pedicle drilling in the spine

Jorge Andrés Pérez Velásquez, François Teyssere, Thibault Chandanson, Quentin Grimal, Brahim Tamadazte

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

Abstract

Purpose: Pedicle screw placement is technically demanding in scoliosis treatment. High precision is required due to limited visibility, anatomical variability, and the risk of complications. Although robotic systems assist CT-based planning and execution, they still rely on ionizing intraoperative imaging and complex registration. This study proposes robotic pedicle drilling with real-time preventive breach detection using electrical bioimpedance sensing.

Methods: We developed a robotic approach combined with a pedicle-drilling tool equipped with a proprioceptive electrical bioimpedance sensor developed by SpineGuard ®^{\circledR}. A real-time detection algorithm was designed to analyze the electrical bioimpedance signal during drilling and identify abrupt changes in conductivity associated with potential breaches towards the spinal canal. The method operates without external devices or sensors.

Results: The ex vivo experiments showed that the proposed method prevented breaches in 100% of the 51 drilling cases. These findings demonstrate the system's ability to detect potentially hazardous events during drilling and to stop the procedure before. The ex vivo experiments demonstrated that the proposed method prevented breaches in all 51 drilling cases.

Conclusions: This work demonstrates the feasibility of robotic pedicle drilling with electrical bioimpedance sensing for real-time breach prevention. Using only the tool signal, the method eliminates the need for external sensing systems and supports safer pedicle screw placement.

Keywords: Surgical robotics, Spine surgery, Pedicle screw placement, Electrical bioimpedance sensing

1 Introduction

Pedicle Screw Placement (PSP) is a surgical procedure used to treat various spinal pathologies, such as scoliosis, spinal trauma, or spinal instability. Drilling the pedicles for screw positioning is a challenging procedure. Failure can lead to injuries, such as major vessels (aorta, vena cava) or the spinal cord [1]. The literature reports a complication rate in the case of screw mispositioning of up to 18% [2]. Indeed, the main risk arises during the pedicle preparation phase, when the surgeon defines a drilling trajectory that is considered optimal and safe for the patient. The surgeon has no means of real-time, in-depth visualization of the vertebra and therefore relies primarily on skill and experience. For example, in scoliosis (Fig. 1 left), which is characterized by severe three-dimensional deformities of the spinal column, identifying the entry point for each pedicle screw is particularly challenging. This often leads to high rates of pedicle screw misplacement, and approximately 7% of these mispositioned screws may result in medial or lateral breaches (Fig. 1, right). Lateral breaches can damage vascular structures, causing significant bleeding and potentially life-threatening complications [3]. When medial breaches occur into the spinal canal, they can expose the spinal cord to risk and lead to neurological consequences ranging from paraparesis to complete paralysis [4]. The literature reports that most breaches are classified as "A" or "B" under the Gertzbein and Robbins classification [5]. In this well-established classification, the position of a transpedicular screw is graded from "A" to "E" based on the extent of pedicle cortex breach: grade "A" corresponds to a fully intrapedicular position without cortical violation, grade "B" to a breach of less than 2 mm, grade "C" to a breach of 2–4 mm, grade "D" to a breach of 4–6 mm, and grade "E" to a breach greater than 6 mm or a screw located outside the pedicle.

To improve the accuracy of pedicle screw placement and prevent breaches, various techniques have been proposed in the literature. First, patient-specific 3D-printed guides generated from preoperative imaging have been introduced as templates to assist PSP procedures [6]. Several studies also report the use of different non-visual exteroceptive or proprioceptive sensors, including vibroacoustic, accelerometric, torque, and force sensors [7–9]. However, most solutions rely on image-based navigation, using fluoroscopy or CT scans.

Furthermore, robotic assistance in spinal surgery was first introduced in 2004, marking a major advancement in the field, which primarily uses navigation systems to optimize pedicle screw placement [10]. Most of the available robotic systems are integrated with stereotactic navigation systems [11]. The SpineAssist system, developed by

Figure

Fig. 1 Left: a typical scoliose deformation. Right: different scenarios for screw insertion into the vertebrae: (1) leading to medial breach, (2) leading to lateral breach, and (3) well-positioned pedicle screw.

Mazor Robotics, was the first robotic system for spinal surgery to receive FDA (Food and Drug Administration) approval. Over the years, significant advances have been achieved, with the latest generation, Mazor X ®^{\textregistered}, incorporating a camera to assess the surgical environment and improve accuracy [12]. Another widely used robotic system is ROSA Spine ®^{\textregistered}, which operates according to similar shared-control principles [13]. The most recent system, ExcelsiusGPS ®^{\textregistered}, integrates real-time intraoperative imaging and enables screw placement without the use of guidewires [14].

Although navigation and robotic systems have improved spinal surgery, they remain limited by usability, cost, and accuracy. Respiratory motion and small patient movements during pedicle preparation can introduce substantial registration errors [15][15]. Moreover, these systems improve trajectory planning but do not fully prevent pedicle breaches. To address this limitation, we developed a robotic platform for pedicle screw placement that integrates a SpineGuard ®^{\textregistered} bioelectrical conductivity sensor, hereafter referred to as the DSG sensor, which is already used clinically [16,17][16, 17]. A threaded drill bit with a pyramidal sensorised tip operating at 25 Hz was mounted on a custom motorized drilling system attached to Kurobot's end effector.

A real-time algorithm uses adaptive conductivity thresholds and a dual-alert mechanism to stop drilling before a breach occurs. The system was evaluated ex vivo on fresh pig vertebrae through 51 drillings, achieving preventive detection in all trials.

The mean error remained within [−1,+1.4][-1, +1.4] mm, and 100% of drillings were graded "A" or "B" according to the Gertzbein and Robbins classification.