arXiv:2610.01452cs.CVOct 2026

Manuscript Offprint

arXiv:2610.01452

Uncertainty-Guided Handshake: Efficient Human-in-the-Loop Refinement for Surgical-Grade Glioma Segmentation

Samuel Hart, Ahmad Yahya, Ahmed Karam Eldaly

Open access. PaperVantage converted this article to text, so layout and figures may differ from the original.

ABSTRACT

While state-of-the-art automated models for medical image segmentation achieve high mean performance, they frequently suffer from localized, catastrophic failures that preclude safe clinical deployment, particularly in neuro-oncology. Interactive segmentation frameworks mitigate this by incorporating human oversight, but traditionally impose prohibitive cognitive and temporal workloads by requiring clinicians to manually search for errors. In this project, we present an efficient, Hybrid Structural-Aleatoric Human-in-the-Loop framework for glioma segmentation that bridges the gap between automated baseline performance and surgical-grade precision, achieving sub-2.0 mm HD95 on curated benchmarks while providing safety-net routing for structural failures across real-world clinical data. By extracting voxel-wise Test-Time Augmentation (TTA) uncertainty and applying hierarchical topological filtering, our method proactively isolates high-risk structural anomalies. We comprehensively evaluated our approach on a challenging out-of-distribution clinical stress-test cohort (N = 362). Operating under a simulated Human Oracle, the framework improved the Whole Tumor (WT) Dice score from 0.891 to 0.914 and reduced the 95th percentile Hausdorff Distance (HD95) from 5.82 mm to 4.76 mm. Critically for surgical safety, the system rescued severe boundary failures in the Tumor Core, reducing mean HD95 from 17.96 mm to 14.83 mm (improving absolute TC Dice to 0.356). These spatial rescues were achieved while demanding a median interactive workload of just 11.3% of the target volume. Acknowledging this as a simulated upper bound lacking real-world cognitive friction, the framework nevertheless demonstrates a highly Pareto-efficient pathway for safely deploying clinical AI.

Introduction

The accurate segmentation of gliomas—the most prevalent, heterogeneous, and infiltrative primary malignant brain tumors—is a critical prerequisite for successful neuro-oncological workflows, including stereotactic surgical planning, radiotherapy target contouring, and longitudinal disease monitoring 1,2^{1,2}. Historically, this process relied entirely on manual volumetric delineation by expert neuroradiologists, a paradigm that introduces extreme intra-observer and inter-observer variability while acting as a severe operational bottleneck in clinical throughput 1,3^{1,3}. To standardize and scale the generation of high-fidelity boundaries, early multi-institutional data curation initiatives focused on establishing annotated digital repositories of glioblastoma and lower-grade glioma variants 4^{4}. Despite these foundational curation efforts, manual annotation remains unviable for rapid intraoperative decision-making, necessitating automated solutions capable of characterizing irregular structural boundaries under tight clinical time constraints 3^{3}.

The landscape of clinical data processing transformed significantly with the deployment of deep learning frameworks across health and medicine 5^{5}. Fully automated medical image segmentation has since been dominated by state-of-the-art deep convolutional networks, most notably self-configuring architectures such as nnU-Net, which automate preprocessing, patch sizes, and normalization strategies for specific imaging modalities 6^{6}. These systems have effectively saturated academic leaderboards, including the regular iterations of the Brain Tumor Segmentation (BraTS) challenges, consistently achieving high global volumetric overlap metrics with Dice Similarity Coefficients (DSC) exceeding 0.90 7,8^{7,8}. Furthermore, these deep architectures have enabled automated quantitative tumor response assessments in multi-centre, retrospective clinical trials, establishing a high baseline for computerized volume tracking 9^{9}.

However, when translated from controlled benchmarks into real-world clinical environments, this global mathematical average frequently masks localized, catastrophic boundary failures 10^{10}. When exposed to heterogeneous, out-of-distribution (OOD) clinical cohorts marked by varying scanner field strengths, gradient non-linearities, and patient motion artifacts, deterministic models exhibit significant degradation 11^{11}. Because standard networks optimize for voxel-wise cross-entropy across an entire population, they are prone to generating severe spatial failures on edge cases. These manifest as massive false-positive structural hallucinations in completely healthy tissue or the total omission of multifocal satellite lesions and diffuse infiltrative margins. In high-stakes neurosurgery, where preserving millimeters of eloquent cortical tissue directly dictates functional patient outcomes, this lack of worst-case topological reliability fundamentally precludes safe, autonomous clinical deployment, creating an urgent requirement for algorithmic explainability and validation systems 2,12^{2,12}.

The pragmatic bridge between fully automated inference and absolute clinical safety is Human-in-the-Loop (HITL) refinement, often formalized as Interactive Medical Image Segmentation (IMIS) 3,13^{3,13}. Traditional IMIS frameworks embed expert human oversight directly into the deep learning pipeline, allowing users to update segmentation predictions iteratively using click-based inputs, bounding boxes, or geodesic scribbles 14,15^{14,15}. Recent foundation models have expanded this paradigm by adapting zero-shot vision models to medical targets through prompt engineering and localized interactive masks 16^{16}. While theoretically appealing, conventional interactive segmentation tools impose prohibitive cognitive and temporal workloads on medical professionals, severely limiting clinical throughput. Standard software suites treat the clinician as a passive visual auditor, forcing them to manually scroll through complex 3D multi-parametric Magnetic Resonance Imaging (mpMRI) volumes slice-by-slice within standard labeling toolkits to visually hunt for localized algorithmic errors before initiating corrections 3,17^{3,17}. This unstructured manual error-hunting process is highly susceptible to human fatigue and oversight, often requiring substantial active interaction time per patient, which largely negates the operational throughput advantages of introducing an automated baseline 3^{3}. Active learning methodologies have attempted to selectively request annotations on unlabelled sets to minimize full-slice editing, yet they continue to face operational barriers when deployed on dense 3D structural volumes 18^{18}. To achieve genuine clinical utility, an interactive framework must shift from a passive tool to an active triage system that proactively guides the clinician's attention directly to regions of peak spatial risk.

To bridge this operational gap, recent research has turned toward Uncertainty Quantification (UQ) as an objective mechanism to identify model ignorance and boundary ambiguity 11^{11}. By utilizing Bayesian approximations, Monte Carlo (MC) Dropout, or Deep Ensembles, modern architectures can generate voxel-wise predictive entropy heatmaps along with standard binary segmentations 19,20^{19,20}. Prior work has shown that predictive uncertainty can support automated quality control, estimating segmentation reliability without ground truth at the scan or structure level 21,22^{21,22}. However, within the existing literature, uncertainty maps are predominantly treated as passive visual overlays, limiting their utility in rapid triage tasks 23^{23}. Clinicians are presented with an informational heatmap but lack an automated, geometric pipeline to translate this spatial variance into actionable modifications 11^{11}. Furthermore, standard 3D Deep Ensembles multiply memory and inference cost by the ensemble size 24^{24}, a latency incompatible with inter-case clinical turnaround, while single-pass alternatives such as Evidential Deep Learning (EDL) 25^{25} derive uncertainty from a heuristic output parameterization whose calibration is not guaranteed under distribution shift 11^{11}—precisely the regime where reliable triage matters most. Test-Time Augmentation (TTA) offers a middle path: genuine predictive sampling from a single trained model, avoiding both the ensemble’s overhead and the single-pass heuristic. Recent approaches have attempted to mitigate this by formulating uncertainty-guided incremental interaction frameworks using sparse Gaussian processes, yet translating raw voxel variance into physical anatomical boundaries remains an open challenge 26^{26}. Consequently, a significant research gap remains in developing an integrated, model-agnostic framework capable of translating raw predictive uncertainty into highly constrained, physically snapped geometric target boundaries optimized for fast clinical verification 3^{3}.

To address these interconnected limitations, we propose the Uncertainty-Guided Human-in-the-Loop (UG-HITL) framework, an active algorithmic triage system designed to transition the neuro-radiologist from a passive error-searcher to an efficient spatial verifier. By extracting high-fidelity epistemic and aleatoric spatial variance 27^{27} through native Test-Time Augmentation (TTA) 20^{20} and applying hierarchical topological rules, our framework acts as a localized spatial “tripwire.” Unlike topology-preserving segmentation methods, which inject connectivity constraints into the training loss 28−30^{28-30}, our rules operate post-hoc at inference over frozen network outputs, functioning as a triage filter rather than a learned shape prior and requiring no retraining or architectural modification. Rather than presenting the clinician with un-contoured volumes or raw entropy maps, the system isolates high-risk structural anomalies and constructs tightly bound, gradient-snapped geometric proposals—termed the “Handshake”—routing only severely failing tumor sub-regions to the clinician for immediate, targeted manual refinement.

The primary contributions of this work are therefore threefold:

  1. We introduce an active, decoupled, backbone-agnostic UG-HITL refinement framework that leverages spatial uncertainty to guide clinician interaction dynamically, reducing the required visual search space down to a tight, localised boundary mask, and we characterize its refinement behavior across three distinct uncertainty estimators (MC Dropout, deep ensemble, and evidential).

  2. We demonstrate that our framework achieves surgical-grade precision (defined as a 95th percentile Hausdorff Distance of <2.0 mm<2.0\ mm) 31^{31} on controlled benchmarks, while establishing a mathematically stable TTA-based inference pipeline

2/12 optimised to securely intercept severe out-of-distribution hallucinations without destabilizing background space.

  1. We quantify the clinical interaction cost across an external, independent cohort (N = 362), proving that critical spatial rescues can be achieved while strictly capping the clinician’s required review workload to a maximum of 25% of the target tumor volume for standard cases, while securely routing total baseline failures to full manual review.