Physics-Constrained Deep Learning Model for Contactless Blood Pressure Monitoring from Triaxial Bodyseismography

Yuanyuan Zhang, Yida Zhang, Jiahui Li, Yuyan Wu, Fei Dou, Xiao Yin, Zhenlin An, Hae Young Noh, Wenzhan Song

Yuanyuan Zhang, Yida Zhang, Jiahui Li, Yuyan Wu, Fei Dou, Xiao Yin, Zhenlin An, Hae Young Noh, Wenzhan Song, Senior Member, IEEE

Abstract—Ballistocardiography (BCG) is promising for unobtrusive long-term blood pressure (BP) monitoring in laboratory settings, but traditional BCG signals are vulnerable to the variations in body-bed interaction with shifted fiducial points in temporal or amplitude axis, and BP varies with personal hemodynamic changes, causing misaligned representations that affect model generalizability and robustness. In this work, we propose a non-invasive BP estimation framework, Phy-BP, based on triaxial bodyseismography (BSG) as an extension of BCG. Firstly, an adaptive quality-control algorithm is designed to select BSG segments enriched with cardiogenic components by jointly considering neighboring beat patterns and universal cardiogenic templates. Furthermore, a physical model is established to describe 3D wave propagation in the body-bed system and is subsequently embedded into the deep learning model to characterize the intrinsic coupling among triaxial BSG signals driven by a single cardiogenic excitation. Thus, multi-axis features are aligned during model training, improving robustness against distortions in real scenarios. Experiments on a 162-hour hospital dataset collected from 21 subjects reveal that the proposed Phy-BP can dynamically filter out low-quality measurements, and the deep learning model training is constrained by physical consistency across different axes to provide faithful BP monitoring, especially when training samples are limited.

Index Terms—Bodyseismography, Ballistocardiography, Blood Pressure Monitoring, Physics-informed Deep Learning

I. INTRODUCTION

Accurate and long-term blood pressure (BP) monitoring is important for capturing nocturnal and longitudinal hemodynamic variations in both home and clinical settings, supporting sleep-related BP assessment, early cardiovascular risk detection and routine overnight monitoring [1], [2]. Although invasive arterial blood pressure (ABP) monitoring remains the gold standard for continuous BP measurement, it requires catheterization and is associated with increased clinical burden and risks such as infection and thrombosis [3], [4]. As a non-invasive alternative, cuff-based devices are widely used for intermittent BP measurement, but their accuracy can be affected by factors such as device algorithms, cuff size, and arm position, while frequent cuff inflation may also cause discomfort, sleep disturbance, and reduced compliance during long-term use [5], [6].

In recent decades, extensive efforts have been devoted to cuffless BP monitoring using non-invasive physiological signals related to cardiovascular function. For instance, photoplethysmography (PPG) and seismocardiography (SCG) have been used to estimate BP through pulse transit time (PTT) or cardiac activity intensity [5], [7]–[11]. However, these contact-based approaches require sustained skin contact and are therefore unsuitable for patients with skin damage or for long-term/overnight monitoring [12]. To achieve more unobtrusive monitoring, camera- and radar-based methods have been explored for remote physiological sensing, but camera systems raise privacy concerns, while current radar methods still suffer from limited SNR and short monitoring range (e.g., within 0.7 meters) under practical conditions [13]–[17].

Compared with aforementioned solutions, bed-mounted vibration sensors can detect the minor displacement caused by heartbeat and body recoil along the foot-head axis (i.e., ballistocardiography (BCG)) [2], [18], [19], as shown in Fig. 1(a). Although the BCG measurements are inherently tied to the bed or other transmission media (e.g., chair [20]), this fixed coupling can help maintain relatively stable signal quality and support continuous overnight monitoring [21]. After decades of development, high-quality BCG can be measured using various sensors (e.g., fiber-optic sensor [2], electromagnetic films [22], seismic sensor [21]), and numerous studies have demonstrated that the BCG signals contain cardiovascular information for downstream applications, such as sleep status monitoring [23] and arrhythmia detection [24], but the BP estimation requires fine-grained cardiac features other than respiration and heart rates.

Existing studies on BCG-based BP estimation can be categorized into two paradigms. The first paradigm requires simultaneous collection of other vital signs (e.g., ECG) to estimate PTT first, and is not purely contactless [20]. In contrast, recent research leverages deep learning models to learn BP-related representations directly from raw BCG waveforms due to the powerful non-linear mapping abilities [25], [26], thereby alleviating reliance on other modalities [1], [2], [22]. However, the existing work shares some major limitations:

This research has been approved by the Institutional Review Board of University of Georgia with application number No.PROJECT00010221, and is also approved by Medical Ethics Committee of Yixing People's Hospital under application No.LS2024-KE-084-02.

Yuanyuan Zhang, Yida Zhang and Wenzhan Song are with the School of Electrical and Computer Engineering, University of Georgia, Athens, 30603, United States, (email: yz52357@uga.edu, wsong@uga.edu).

Jiahui Li, Fei Dou and Zhenlin An are with the School of Computing, University of Georgia, Athens, 30603, United States.

Yuyan Wu and Hae Young Noh are with the Department of Civil and Environmental Engineering, Stanford University, Stanford, CA, 94305, United States.

Xiao Yin is with the Yixing People's Hospital, Wuxi 214200, China. Corresponding author: Wenzhan Song

Figure

(a) SBP = 124, DBP = 77

(b) SBP = 125, DBP = 73

(c) SBP = 122, DBP = 80

Fig. 1. Illustration of BSG sensors placement and signals under similar BP: (a) Standard BSG signal with the majority of features shown in Y-axis; (b) Energy leakage to X-axis with no observable features in Y-axis; (c) Distorted BSG signal due to soft/thick mattress with attenuated and shifted fiducial peaks.

• Inaccurate BP ground truth measurement: Most BP estimation studies use intermittent BP readings from oscillometric devices as ground truth for convenience, while such measurements cannot capture beat-to-beat BP dynamics and thus cannot provide continuous hemodynamic monitoring as a potential clinical reference [2], [5]. In addition, the accuracy of cuff-based measurements is highly affected by cuff size/placement, arm posture, and physiological fluctuations during repeated inflation, with absolute measurement errors reaching up to 28.3 and 18.5 mmHg for systolic and diastolic BP (SBP and DBP), respectively [27], [28].

• Unfaithful cardiac features in 1D BCG: Standard 1D BCG signals capture the body recoil along the Y-axis (foot-head direction), and physiological experiments have revealed the correlation between BCG fiducial points (i.e., I, J, K peaks) and cardiac force [19]. However, according to our measurement, 1D BCG cannot faithfully capture the complete bed response induced by cardiac activities, and a considerable portion of the vibration energy may leak into the X- and Z-axes under different postures, as shown in Fig. 1(b). In addition, BCG signals are vulnerable to the change of body-bed interactions affected by mattress and body weight, with attenuated features shown in Fig. 1(c).

• Reliance on purely data-driven methods: Although many studies have realized the correlations between BCG fiducial points and cardiac force [2], [19], [20], such relations are not explicitly modeled and extrapolated to BP estimation in terms of cardiogenic excitation and wave propagation theory. Most existing work uses a deep learning model as a black box to learn a domain transformation from numerous BCG-BP pairs, while uncertainties in BP ground truth and BCG morphology lead to misaligned representations during feature extraction, resulting in well-trained models that lack generalizability and robustness [29].

In this study, we leverage triaxial bodyseismography (BSG) [30] measured by seismic sensors (Fig. 1), as an extension of the 1D BCG signal for BP estimation to faithfully reveal the cardiac features. In addition, the invasive ABP waveform is used as a golden-standard reference for the model training and evaluation, along with simultaneously collected ECG and PPG signals for a fair and comprehensive comparison. At last, we model the 3D wave propagation and propose a BSG-based BP estimation framework to properly extract and align triaxial BSG features. The contributions of this work are listed as follows:

  • To the best of our knowledge, this is the first research that explicitly characterizes how cardiogenic excitation propagates through the body-bed coupling system. We establish a physical model based on wave propagation theory to interpret how the measured triaxial BSG signals are jointly shaped by cardiogenic driving forces and the dynamic response of the body-bed system.

  • A BSG-based BP estimation framework, Phy-BP, is proposed with an adaptive quality-control module that filters out low-SNR signals by considering both neighboring beat patterns and universal cardiogenic templates, ensuring enriched and faithful cardiogenic features for further BP estimation.

  • Phy-BP also contains a deep learning model that embeds the proposed physical model as a constraint to characterize the intrinsic coupling among triaxial BSG signals induced by a single cardiogenic excitation, ensuring robust feature extraction and alignment across three axes.

  • The proposed Phy-BP was validated on a 162-hour hospital dataset collected from 21 patients, using invasive ABP measurements as the reliable reference. Extensive experiments demonstrate that Phy-BP outperforms existing BCG- and PPG/ECG-based BP estimation methods, and the results satisfy the requirements of the Association for the Advancement of Medical Instrumentation (AAMI).

The rest of the paper is organized as follows. Section II provides the hemodynamic background and challenges for BSG-based BP monitoring. Section III elaborates on the physical and deep learning model designs, with the experimental settings and dataset overview presented in Section IV. All the results are illustrated and evaluated in Section V, and Section VI concludes this research.