TRACE: Two-Stage Detector-Response Estimation With Angular Cosine Expansion for Ring Artifact Correction in Photon-Counting CT

Jigang Duan, Heran Wang, Ligen Shi, Zheng Sun, Ping Yang, Xing Zhao

Jigang Duan, Heran Wang, Ligen Shi, Zheng Sun, Ping Yang, and Xing Zhao

Abstract—Detector response nonuniformity introduces systematic projection errors and ring artifacts in photon-counting detector computed tomography (PCD-CT). In measured PCD-CT data, residual stripe amplitudes vary slowly with projection angle, which fixed-bias models cannot adequately capture. We propose TRACE, a two-stage unsupervised sinogram decomposition method for estimating and correcting these response-related errors. TRACE represents stripes as a fixed bias plus low-order discrete cosine transform (DCT) components, using a small number of coefficients to describe angular variations at each detector element. A learnable analysis–synthesis architecture represents the ideal projections, while two-stage optimization separates them from fixed and then dynamic stripes. An angular-gradient soft orthogonality constraint suppresses correlated variations within the shared DCT gradient subspace, reducing the leakage of object structures into the artifact estimate. All parameters are optimized directly on the measured sinogram without paired training data. Experiments on measured QRM mouse phantom and porcine trotter data show that TRACE suppresses ring artifacts and improves image uniformity while preserving edge sharpness, soft-tissue texture, and trabecular detail.

Index Terms—Photon-counting CT, detector response nonuniformity, ring artifact correction, sinogram decomposition, discrete cosine transform, unsupervised learning

I. INTRODUCTION

Response nonuniformity among detector elements introduces systematic errors into computed tomography (CT) measurements. These errors appear as stripes extending along the angular direction in sinograms and form rings around the rotation center after reconstruction [1], [2]. Unlike conventional energy-integrating detector CT, photon-counting detector CT (PCD-CT) registers individual X-ray-induced electrical pulses and discriminates their amplitudes using energy thresholds. Differences in counting response between detector elements can depend on energy threshold and incident flux, so fixed calibration parameters may not fully compensate for response nonuniformity throughout a scan. Residual response errors can produce stripes that vary with energy window and projection angle [3]–[5]. These artifacts impair image uniformity, distort reconstructed attenuation values, and obscure low-contrast structures. Correcting residual response errors while preserving object information is therefore important for reliable PCD-CT measurements.

References (41)