A comparison of similarity measures for use in 2-D-3-D medical image registration

Graeme Penney, Jürgen Weese, John A. Little, Paul Desmedt, David Hill, David J. Hawkes

Year
1998
Venue
IEEE Transactions on Medical Imaging
References
21
Cited by
897

Details and counts from OpenAlex.

Abstract

A comparison of six similarity measures for use in intensity-based two-dimensional-three-dimensional (2-D-3-D) image registration is presented. The accuracy of the similarity measures are compared to a "gold-standard" registration which has been accurately calculated using fiducial markers. The similarity measures are used to register a computed tomography (CT) scan of a spine phantom to a fluoroscopy image of the phantom. The registration is carried out within a region-of-interest in the fluoroscopy image which is user defined to contain a single vertebra. Many of the problems involved in this type of registration are caused by features which were not modeled by a phantom image alone. More realistic "gold-standard" data sets were simulated using the phantom image with clinical image features overlaid. Results show that the introduction of soft-tissue structures and interventional instruments into the phantom image can have a large effect on the performance of some similarity measures previously applied to 2-D-3-D image registration. Two measures were able to register accurately and robustly even when soft-tissue structures and interventional instruments were present as differences between the images. These measures were pattern intensity and gradient difference. Their registration accuracy, for all the rigid-body parameters except for the source to film translation, was within a root-mean-square (rms) error of 0.54 mm or degrees to the "gold-standard" values. No failures occurred while registering using these measures.