Paper record

W2102012860

Three-dimensional display in nuclear medicine

Jerold W. Wallis, T.R. Miller, C A Lerner, E.C. Kleerup

Year
1989
Venue
IEEE Transactions on Medical Imaging
References
16
Cited by
201

Details and counts from OpenAlex.

Abstract

Several surface and volume rendering techniques are compared using nuclear medicine data including several new methods developed by the authors specifically for scintigraphic data. The techniques examined are summed projection, thresholded projection, threshold-based surface illumination, volumetric compositing, maximum-activity projection, sun-weighted maximum-activity projection, and variable attenuation. The advantages and disadvantages of each method are discussed in relation to the goals of three-dimensional display, which are defined herein. Selected images are shown to illustrate the usefulness of the methods.